WITE32 Release Notes Version 3.20
The Working Zone frame allows you to specify a working zone to be erased and the position where the
service track will be written.
The Track Number text box Specifies the track number, where service track will be written.
The Zone Width text box Specifies the width of the working zone of the test. This zone will be
erased and then used for the track detection stroke movements.
The Track Detection Parameters frame allows you to specify the track profile parameters needed to
calculate the precise track radius after the track detection stroke movements.
The From text box Specifies the starting offset of the track profile. The absolute value
cannot exceed the maximum offset.
The To text box Specifies the final offset of the track profile. The absolute value
cannot exceed the maximum offset.
The Step text box Specifies the step by which the head is moved between offsets to
build the track profile. This value cannot be smaller than the device
step.
The Save to DB check box, if enabled, saves the test results to a database.
The Write Service Track button allows you to write the service track, which will be used for alignment.
The Save button saves all modified parameters to a database.
The Reset button restores all the parameters, which were early saved to the database.
The Default button restores the default values for all parameters.
The Run button starts the test execution.
3.3.3 Lock Motors Mode
The Lock Motors mode is implemented for V2002 spinstand as a positioning mode for troubleshooting. In this
mode the linear motors are locked, so the spinstand does not move the X and Y stages during any spinstand
operations. The micro positioning system still can move the head, when the spinstand is in this mode.
You can use the Lock Motors mode, for example, if you need to position the head over the media manually, start
the device, perform measurements, and stop the device without moving the stages. This feature is also helpful
when you cannot perform XY Alignment and/or Start Device for some reason. It may occur when there is no
signal from the head or the signal is weak.
Guzik Technical Enterprises 15