Operation Manual
FULL AUTO REF
Before using the instrument, be sure to read this manual thoroughly.
Keep the manual in easy-to-access place.
Seitenansicht 0
1 2 3 4 5 6 ... 69 70

Inhaltsverzeichnis

Seite 1 - Operation Manual

Operation ManualFULL AUTO REFBefore using the instrument, be sure to read this manual thoroughly.Keep the manual in easy-to-access place.

Seite 2 - PLEASE NOTE

(8)Safety InformationWarning LabelThe unit weighs 21 kg –it may tip over if propercare is not taken.Rating Plate and Warning LabelThe R-F10 has a rati

Seite 3 - Safety Information

ContentsSafety Information ... (1)1. Overview ...

Seite 4 - ISO15004

7. Daily Inspection and Maintenance ...357.1 Inspection...

Seite 5

– 1 –1. OverviewThe Canon Full Auto Ref R-F10 is for performing refractometry.The main feature of the R-F10 is that by just displaying the examinee’s

Seite 6 - General Safety Information

– 2 –2. Notes for Using the InstrumentBefore Use(1) Sudden heating of the room in cold areas will cause condensation to form on theprotective glass in

Seite 7

– 3 –3. Description3.1 Main Unit 1 Measurement HeadUnit that performs measurement. 2 Height Adjustment MarkAlign the height of the examinee’s eye wi

Seite 8 - Handling

– 4 –3.1 Main Unit 7 Face RestPlace the examinee’s face against this rest. 8 PrinterPrints measurements. 9 Rating PlateName of the product, rated

Seite 9 - Maintenance and Inspection

– 5 –3.2 Connectors under the Main Unit3.2 Connectors under the Main UnitNOTE: Do not remove the cover for the RS232C connector.Please contact Canon

Seite 10 - Mirar el manual

– 6 –3.3 Operation Panel3.3 Operation Panel 1 DISP SwitchPress this switch in order to enter DISPLAYmode, where you can see measurement datastored i

Seite 11 - Contents

– 7 –3.3 Operation Panel 8 SET SwitchPress this switch in order to enter SET mode,where you can perform various settings con-cerning measurements. 9

Seite 12

PLEASE NOTE1. The user is responsible for the use and maintenance of the product.We suggest that a member of the user’s staff be designated as being i

Seite 13 - 1. Overview

– 8 –3.3 Operation PanelDISP switchIOL switchVD switchR/L switchTrackballPRINT switchSET switchMANU. switchCHIN RESTswitchesRollerSTART switchAUTO sw

Seite 14 - During Installation

– 9 –4. MeasurementsWhen any operation has been interrupted for more than 3 minutes with the power turnedON, power-saving system of the R-F10 operates

Seite 15 - 3. Description

– 10 –4.1 PreparationWhen the checks are completed,FULL AUTO mode will be selectedand the measurement head will moveand stop at the right eye side.NO

Seite 16 - 3.1 Main Unit

– 11 –4.1 Preparation3Adjust the height of the examinee’s eye(1) Have the examinee sit and place his/her chin and forehead against the chin restand f

Seite 17

– 12 –4.2 Measurement in FULL AUTO Mode4.2 Measurement in FULL AUTO ModeThere are two modes of performing measurement; FULL AUTO and MANUAL.FULL AUTO

Seite 18 - 3.3 Operation Panel

– 13 –4.2 Measurement in FULL AUTO Mode1Check that automatic measurement has been chosenIf the words “FULL AUTO” is notdisplayed on the monitor, pres

Seite 19 - 3.3 Operation Panel

– 14 –4.2 Measurement in FULL AUTO Mode4Start automatic measurementPress the START switch.Measurement head moves automati-cally to perform alignment.W

Seite 20

– 15 –4.2 Measurement in FULL AUTO Mode– Continuous manual measurement –When START switch is pressed for about two seconds after the messageis displa

Seite 21 - 4. Measurements

– 16 –4.2 Measurement in FULL AUTO Mode5After the alignment, measurement is performed automaticallyMeasurement is performed as many times as selected

Seite 22 - Check the settings

– 17 –4.2 Measurement in FULL AUTO Mode6Move measurement head toward the other eyeIf “R–L MEASURE” is set to “ON” in SET mode (→ section 6.1), the me

Seite 23 - NOITUAC!

(1)Safety InformationRegulationsFor U. S. A.This instrument is a CLASS I EQUIPMENT and TYPE B APPLIED PARTS according toUL2601-1.Do not make any chang

Seite 24

– 18 –4.2 Measurement in FULL AUTO Mode9Measurement head returns to initial positionWhen printing is completed, the measurement head returns to its in

Seite 25 - FULL AUTO” is not

– 19 –4.3 Measurement in MANUAL Mode4.3 Measurement in MANUAL ModePerform measurement in MANUAL mode when the pupil is eccentric, or when measuring a

Seite 26 - Start automatic measurement

– 20 –4.3 Measurement in MANUAL Mode4Check that the alignment ring is not covered by eyelashesLooking at the monitor, check that the inner alignment

Seite 27

– 21 –4.3 Measurement in MANUAL Mode6Perform focusingTurn the roller back and forth to position the three bright dots vertically between thevertical

Seite 28

– 22 –4.3 Measurement in MANUAL Mode8Measure the other eyePress the R/L switch to move themeasurement head to the other side.Measure the left eye in

Seite 29 - The result is printed out

– 23 –5. Measurements Stored in Memory [DISPLAY Mode]In DISPLAY mode, you can see the results (maximum of 10 data for each eye) of measure-ments store

Seite 30 - Example of Printout

– 24 –6. Various Settings [SET Mode]Various settings concerning measurement, printing and data transfer can be performed in theSET mode.There are thre

Seite 31 - Choose manual measurement

– 25 –6.1 Settings for Measurement (Page: 1/3)6.1 Settings for Measurement (Page: 1/3)6.1.1 Basic Operation[How to Select the Item]Turn the roller to

Seite 32

– 26 –6.1 Settings for Measurement (Page: 1/3)INC Increment of sphere and cylinder “0.12, 0.25”COUNT Whether to display and print the serial number“ON

Seite 33

– 27 –6.1 Settings for Measurement (Page: 1/3)“OFF”: Only one eye will be measured. When this is selected,printing must be done by pressing the PRINT

Seite 34 - Print/transfer the result

(2)Safety InformationISO15004This report provides information about the hazard to the examinee’s eyes in compliance withISO15004 (1997).1. The spectra

Seite 35

– 28 –6.2 Settings for Printing and Transfer (Page: 2/3)6.2 Settings for Printing and Transfer (Page: 2/3)6.2.1 Basic Operation[How to Select the Item

Seite 36 - How to Enter SET Mode

– 29 –6.2 Settings for Printing and Transfer (Page: 2/3)6.2.2 ItemsPRINT Whether to print the result“ON”: Results are printed with the internal printe

Seite 37 - 6.1.2 Items

– 30 –6.2 Settings for Printing and Transfer (Page: 2/3)[FMT] Transfer formatNOTE: For details concerning format, please contact Canonrepresentative o

Seite 38

– 31 –6.3 Entering Message for Internal Printer (Page: 3/3)6.3 Entering Message for Internal Printer (Page: 3/3)Enter a message to be printed out with

Seite 39

– 32 –6.3 Entering Message for Internal Printer (Page: 3/3)6.3.2 How to Enter the Characters1Move the position cursor to the required positionPress IO

Seite 40 - 6.2.1 Basic Operation

– 33 –6.3 Entering Message for Internal Printer (Page: 3/3)6.3.4 How to DeleteIn order to delete single character and shift all the characters onthe r

Seite 41 - 6.2.2 Items

– 34 –6.3 Entering Message for Internal Printer (Page: 3/3)2Select the line “CHAR CLEAR / LINE CLEAR / ALL CLEAR”Press PRINT switch.3Select “LINE CLEA

Seite 42

– 35 –7. Daily Inspection and Maintenance7.1 InspectionIn order to ensure that the instrument is used safely and normally, please be sure to inspectth

Seite 43 - ••••••••••••••••••

– 36 –7.1 Inspection7.1.1 Before Turning ON the PowerIt is recommended that a record of the inspection be kept by copying this and the next page,or m

Seite 44

– 37 –7.1 Inspection7.1.2 After Turning ON the PowernoitcepsnItluseRydemeRetaD/etaD/etaD/tinuniaMYDAERehttahtkcehCdnasthgilpmalnoyalpsidtnemerusaem.l

Seite 45 - In order to delete a line

(3)Safety InformationIf the value of radiance was 40 [mW/cm2/sr], 6 minutes would be needed to reach the rec-ommended limit.That is, the retinal expos

Seite 46

– 38 –7.2 Before Calling a Service Person7.2 Before Calling a Service PersonWhenever a problem occurs or message appears on the monitor, take the cou

Seite 47 - 7.1 Inspection

– 39 –7.2 Before Calling a Service Person7.2.2 If Message Such as Following Appears on the MonitorIf message does not disappear even after taking the

Seite 48 - GNINRAW!

– 40 –7.2 Before Calling a Service PersonyalpsiD/egasseMesuaCydemeRgniruDcitamotuatnemngilaRESUYBDELLECNAC]01[dellecnacneebsahtnemngilAsahrenimaxeeht

Seite 49 - 7.1 Inspection

– 41 –7.2 Before Calling a Service Person↓yalpsiD/egasseMesuaCydemeRgniruD-erusaemtnemRESUYBDELLECNAC]31[launamsuounitnoCneebsahtnemerusaemehtesuaceb

Seite 50

– 42 –7.2 Before Calling a Service Person* “xxx” indicates a three-digit number.yalpsiD/egasseMesuaCydemeRgniruDgnitnirphtiwlanretniretnirpREPAPTNIRP

Seite 51 - EDOMLAUNAMNIYRT

– 43 –7.3 Cleaning and Disinfection7.3 Cleaning and Disinfection7.3.1 Protective Glass in Measurement WindowClean the protective glass in the measure

Seite 52

– 44 –7.3 Cleaning and Disinfection7.3.2 Forehead Rest7.3.3 Cover, Monitor and RollerClean the cover and monitor if they are dirty.1Turn OFF the powe

Seite 53

– 45 –7.3 Cleaning and Disinfection7.3.4 TrackballClean the trackball if it does not rotate smoothly.(1) Turn the trackball cover in counter-clockwis

Seite 54

– 46 –7.4 Replacement7.4 ReplacementPlease refer to “7.5 Expendable Parts List” for ordering chin rest paper and printing paper.7.4.1 Chin Rest Paper

Seite 55 - 7.3 Cleaning and Disinfection

– 47 –7.4 Replacement7.4.2 Printing PaperReplace the roll of printing paper as soon as possible after the red line appears on the paper.1Open the pri

Seite 56

(4)Safety InformationGeneral Safety InformationFollow the safety instructions in this manual and all warnings and cautions printed on the warninglabel

Seite 57 - 7.3.4 Trackball

– 48 –7.4 Replacement5Load the roll of paperLoad the roll of paper into the printeras illustrated.NOTE: Be sure to check that thedirection of roll is

Seite 58 - 7.4 Replacement

– 49 –7.5 Expendable Parts ListNOTES: 1. If the edge of paper is creased, it may not feed through printer properly. In thiscase, pull out the paper,

Seite 59 - 7.4.2 Printing Paper

– 50 –8. Installation8.1 InstallationNOTES: 1. Do not install the instrument in a location with the conditions listed on page(2).2. Do not remove the

Seite 60 - Turn ON the power

– 51 –8.1 Installation2. Do not lay the instrument on its side with the power turned ON. Otherwise, theinstrument may malfunction.3Connect the power

Seite 61 - 7.5 Expendable Parts List

– 52 –8.2 Precauations when Moving the Instrument8.2 Precautions when Moving the InstrumentNOTE: When the instrument is going to be transported by au

Seite 62 - 8. Installation

– 53 –9. Service InformationRepairIf problem cannot be solved even after taking the measures indicated in sections 7.1 and 7.2,contact Canon represent

Seite 63 - 8.1 Installation

– 54 –10. SpecificationsPurpose: Measurement of refractive power of humaneyeMeasurement:Vertex distance 0, 12, 13.5 mmMeasurable rangeSphere power -30

Seite 64

– 55 –11. ComponentsR-F10 main unit ... 1Power supply cable ...

Seite 65 - 9. Service Information

– 56 –This page has been left intentionally blank.

Seite 67 - 11. Components

(5)Safety InformationInstallation OperationPower SupplyGNINRAW!.deificepsnahtrehtognihtynahtiwtnemurtsniehttcennoctonoD.kcohscirtceleroerifnitluseryam

Seite 68

CANON INC. Medical Equipment Group20-2 Kiyohara-Kogyodanchi, Utsunomiyashi, Tochigiken, JapanTelephone: (81)-28-667-5711CANON MEDICAL SYSTEMS Eye